· 2GBS is a high-performance, high-quality glove box with a sealed purification system that automatically absorb moisture and oxygen molecules and purifies the working environment.

It can form an inert atmosphere of H2O&O2<0.1PPM that satisfy your special laboratory requirements. This economic purification system is designed to satisfy the research and development requirements of customers.





Product model

2GBS

Other types





  • Structure composition and principle
  • Application area
  • Characteristics
  • Specifications
  • Structure composition and principle

    2GBS is composed of a sealed main chamber, a set of antechamber(including big and small antechamber) that can be evacuated, a rotary vane vacuum pump and a display with PLC controller.

    The gas purification system is composed of a single automatic and renewable purification column and a blower to purify and maintain the working environment. It form a inert atmosphere of H2O&O2 ≤ 0.1PPM to satisfy your

     laboratory requirements.



    Application area

    Mainly creating a non-moisture and non-oxygen environment for organic metal chemistry, organic synthesis, 3D printing of new material powders, special anaerobic and hydrophobic

    chemical treatment, biomedical,electronic component processing, solar cell assembly, lithium battery materials and battery assembly experements, catalyst treatment, pharmaceutical

    synthesis, nuclear industry, membrane industry, etc;





    Characteristics

    • SUS304 stainless steel box design, tempered glass or polycarbonate front window (opional);

    • Indicator: H2O&O2 ≤ 0.1PPM

    • Leakage rate ≤ 0.001 vol%/h;

    • The solenoid valve adopts modular design, which reduces leakage rate and is easy to replace

    • A large antachamber with a sliding tray, featuring a unique design for the transition door that is lightweight and easy to open;

    • Inert gas purification system; HEPA high-efficiency filter;

    • SIEMENS microcontroller; SIEMENS operation touch screen, easy to access various functions;

    • Sealed gas circulation, oil-free and vacuum free; All stainless steel gas flow pipes and accessories;

    Data Memo: Automatic Recording System Data;

    • Replacing manual pressure control with foot pedals;

    • Automatic pressure control, the working pressure can be set within the range of+12mbar;




    Specifications


    Main Chamber
    Size1220mm(L)*760mm(W)*900mm(H)  
    Front window

    Panel: tempered glass of 8mm thickniss or polycarbonate material of

     10 mm thickniss (optional)

    Glove port

     Aluminum alloy material or polyformaldehyde material (optional): 

    220mm diameter with O-ring sealed

    GloveInport butyl rubber
    HEPA filtersInlet and outlet filter can eliminate particle size <0.3 μm
    LightingLED lighting is located at the top of the front window
    Leakage range

    Conduct final acceptance testing according to the standard

     of leakage rate < 0.001vol%/h

    Sensor

    Oxygen analyzer 

    (optional zirconia)

    USA AII
    Accuracy ± 0.1PPM


    Electrochemical cells have the advantages of accurate zero point, small drift, 

    less influence of organic 


    solvents on detection results, and low replacement costs 

    (only replacing electrochemical cells). Other

     

    types of probes (zirconia) can also be replaced according to customer needs.


    Moisture analyzerVaisala, Finland
    Accuracy ± 0.1PPM
    Range 0-1000ppm, low drift
    Gas purification system

    Describe


    Automatically remove H2O and O2; Single purification column system; 


    Automatic regeneration (optional high-efficiency dual purification column);


     Sealed gas circulation pipeline;


    VoltageAC 230V/50-60 Hz, 10A or AC 115V/50-60Hz, 20A (optional)
    Gas requirements


    Working gas: N2, Ar, He (purity ≥ 99.999%); 


    Regeneration gas: H2 accounts for 5% -10%, the rest is working gas (Purity ≥ 99.999%)


    Vacuum pump


    Specification: rotary vane vacuum pump, equipped with oil mist filter,

     with pneumatic control;


     Flow rate: 8m3/h (7cfm), vacuum degree<2x10-3mbar (optional dry pump)


    Loop unitOil free high-speed fan air volume: 0-90m3/h
    VavleElectromagnetic high vacuum valve
    Leakage range

    Conduct final acceptance testing according to the standard

     of leakage rate<0.001vol%/h

    Optional selection

    Organic solvent 

    adsorption 

    system


    Material: Stainless steel SUS type 304 with a thickness of 3mm;

    Internal dimensions: 220mm(diameter) x 450mm (H); 

    High quality activated carbon;


    Refrigerator


    Independent control system, integrated on the side of the main chamber;


     temperature -35℃, optional capacity of 18L or 32L;


    Main body made of 304 stainless steel plate, with 5 movable shelves;


    R404 environmental protection refrigerant, imported with original packaging;


    Heating 

    antechamber


    The large antechamber is equipped with a heating system, with a temperature of 200 ℃ 


    and a temperature control of ± 1 ℃